APPLICATIONS
• The solid-state, ultra-high-tech Low Temperature Cabinet (Horizontal/Vertical Models) microcontroller-controlled refrigeration system, free of CFCs, is designed to meet standard applications in a variety of industrial settings, including biochemical, medical, and research laboratories.
• Used for the preservation of red blood plasma, white blood cells, skin, bone, bacteria, semen and biomedical products in small, medium, and large blood centres.
FEATURES
• The outer body is made of mild steel that has been duly prepared and powder coated.
• Inner chamber made of stainless-steel sheets (304 Grade).
• Ergonomically designed.
• The double-walled, insulated outer door has a silicone gasket installed for an airtight closure.
• The outer front double-walled door has a lock and key system, is insulated, and has a silicone gasket installed for an airtight closure that prevents temperature loss.
• The Ultra Low Temperature Cabinet is fitted with Cascade Refrigeration System.
• A state-of-the-art refrigeration system with a cascading system and two compressors.
• Adjustable shelves are convenient to place articles.
• Microprocessor-based temperature controller with indicator. Set Value (SV) and Process Value (PV) LED display allows users to choose from a range of available desired temperatures, from -10°C to -80°C.
• Exceptional visual and auditory alert systems (fault alarms for low or high temperatures, low battery, open doors, obstructed filters and system failure)
• Castor wheel mounted for convenient portability.
• Comes with plug and cord.
• Operating Voltage: 220 Volts AC (50 Hz).
TECHNICAL SPECIFICATIONS
Model No. | ULF201 | ULF401 | ULF501 | ULF701 | ULF901 |
Approximate volume | 232/217 | 478/454 | 634/602 | 791/451 | 949/900 |
Refrigerant Type | R290 / R170 | ||||
Temperature range | -20°C to -80°C | ||||
Temp. controller | By microprocessor based PID digital temperature indicator | ||||
Temp. Accuracy | +1 °C | ||||
Display | LED display for set value and process value | ||||
Temperature Alarm | For high and low temperature variations | ||||
Relay | Solid static electric relay with protective heat | ||||
High Voltage Protection | Through automatic voltage stabilizer (Optional) | ||||
Electric Supply | 220/230V AC, 50/60Hz |
OPTIONAL FEATURES
PLC Based Control Systems | Touch Screen Display with Direct Thermal / Dot Matrix Printers with optional Rs 485 communication portal cable |
LCD controller with data controller | Window based software with inbuilt data recording |
Printing facility | Through built-in mini printer to record temperature |
Temperature recorder | By weekly temperature recorder |
Communication port | Communication port with interface and data cable to download data to your pc. |
Alarms | High or low temperature alarm, battery alarm, door open alarm, filter blocked alarm, system failure alarm |